|
|
|
Categories
|
|
Information
|
|
Featured Product
|
|
|
 |
|
|
There are currently no product reviews.
 ;
Downloaded the manual, reasonably straightforward, pretty much exactly as advertised.
 ;
very helpfull
circuit diagram and sparepart list available
 ;
Very good reproduction (copy) of original manual. Didn't have a parts list, but schematic was completely labeled with parts. Complete instructions on how to adjust mechanical functions of the 8-track deck. Well worth having and at a very reasonable cost.
 ;
It's a full manual. All the parts are in there. I haven't found the problem yett, but I am working on it; hope I can rebuild the part myself. To make it more secure and unbreakable this time. Because the part has failed several times before and costs a lot to let it be repaired.
Thanks so much for this rich illustrated and parted manual.
 ;
I downloaded the document. The manual was complete, well scanned and everything was legible. I could zoom in see what I needed to know. There's not much more that you can ask.
88
Service Manual
Referring to Figures 3-1 and 3-2, perform the following procedure to examine or replace the meter�s fuses: 1. Perform steps 1 through 3 of the battery replacement procedure. 2. Remove the defective fuse by gently prying one end of the fuse loose and sliding the fuse out of the fuse bracket. 3. Install a new fuse of the same size and rating. Make sure the new fuse is centered in the fuse holder. 4. Make sure that the Meter rotary switch and circuit board switch are in the OFF position. 5. Ensure that the gasket remains secured to the case bottom. Then, starting with the two snaps at the display end and finishing at the terminal end, cradle the case top into the gasket. 6. Reinstall the three screws from the case bottom.
3-10. Cleaning
Warning
To avoid electrical shock or damage to the meter, never allow water inside the case. To avoid damaging the meter�s housing, never use solvents on the meter. If the Meter requires cleaning, wipe it with a cloth that is lightly dampened with water or a mild detergent. Do not use aromatic hydrocarbons, chlorinated solvents, or methanolbased fluids on the Meter.
3-11. Input Terminals Water, dirt, or other contamination in the A or mA terminals may activate the Input Alert (beeper) even though test leads are not inserted. Such contamination can sometimes be dislodged by turning the Meter over and gently tapping on the case. However, Fluke recommends the following procedure as a much more effective treatment:
1. Turn the Meter off and remove all test leads. 2. Use a clean swab in each of the four terminals to dislodge and clean out the contamination. 3. Moisten a new swab with a cleaning and oiling agent (such as WD40). Work this swab around in the A and mA input terminals. The oiling agent insulates the terminals from moisture-related shorting. This preventive treatment ensures against future erroneous Input Alerts.
3-12. Function Encoding Potentiometer Perform the following procedure to clean the rotary switch potentiometer, which is screened on the back of the pca:
1. Remove the circuit assembly as described previously in this section under "Removing and Reinstalling the Circuit Assembly". 2. From the back of the circuit assembly, push the switch shaft in (see Figure 3-2), and remove the polymer thick-film (PTF) contact assembly. 3. Push and secure the ptf contact assembly back on the switch shaft.
3-8
|
|
 |
> |
|