|
|
|
Categories
|
|
Information
|
|
Featured Product
|
|
|
 |
|
|
There are currently no product reviews.
 ;
Very quick and easy website to use and fast download of manual, quality of manual is excellent and will be pleased to use this service again in the future, thanks so much!
 ;
It is an very good and clear scanned service manual.
very recommended.
 ;
Easy to order the manual. Good quality and fast delivery.
 ;
The Service Manual for Sansui AU-9500 was very helpfull, in complete and in good printable condition.
Thanks.
 ;
Dear Sir,
Document is original service document of sharp. I had a problem with the door contacts. Fuses where blown. With the manual in a few minuts is was clear what the problem was.
Manual was of great help.
With kind regards,
Martie Verhoeven
The Netherlands.
4-1. Check Items Before Starting Self-diagnostics
3. When the DE-55 board is suspected to be defective . Select an oscilloscope setup of 20 MHz with band-width limit: OFF. . Adjust the brightness of the incoming light until 400 mVp-p is obtained at TP401 and TP402. . The circuit stage up to TP141 must be diagnosed by observing the waveform of the carrier signal and by checking the operating condition with no input signal fed to the infrared transmitter. The circuit stages after that must be diagnosed by observing the demodulated waveform and by checking the operating condition with the color bar signal being input to the infrared transmitter. a) Check that the DC voltages that are supplied from the power supply to the respective blocks are correct or not. � When the DC voltages are normal, the voltages from MA-107/107P board are; TP491: 5 ± 0.5 V, TP391: 9 ± 0.5 V. � When DC voltages are abnormal; Soldering of CN101 is suspected to be defective, or flexible flat cable (FFC) has poor connection or is broken. b) Check DC voltage at L491 and check the signal flow from TP401 to TP404. [RF amplifier] � When the circuit is normal, the waveforms [No. 1 to No. 5] that are shown on page 4-21 can be obtained with no input signal fed to the infrared transmitter. � When the circuit is found abnormal; Resoldering the poorly soldered joint in the circuit block of Ref. 400 to Ref. 409. Replace the transistor as needed. c) Check DC voltage at L492 and check the signal flow from TP421 to TP424. [RF amplifier] � When the circuit is normal, the waveforms [No. 1 to No. 5] that are shown on page 4-21 can be obtained with no input signal fed to the infrared transmitter. � When the circuit is found abnormal; in the circuit block of Ref. 420 to Ref. 439. Replace the transistor as needed. d) Check DC voltage at L491. [ANALOG_5.9 V] � When the circuit is normal; 5.9 ± 0.5 V � When the circuit is found abnormal; Resoldering the poorly soldered joint in the circuit block in which abnormality has occurred. Replace the transistor Q901 or others as needed. e) Check Q110 for correct operation by observing the output waveform at TP101. [BUFFER] � When the circuit is normal, the waveform [No. 6] that is shown on page 4-21 can be obtained with no input signal fed to the infrared transmitter. � When the circuit is found abnormal; Resoldering the poorly soldered joint in the circuit block in which abnormality has occurred. Replace the transistor Q110 or others as needed. f) Check IC101 for correct operation by observing the output waveform at TP102. [LIMITTER] � When the circuit is normal, the waveform [No. 7] that is shown on page 4-22 can be obtained with no input signal fed to the infrared transmitter. � When the circuit is found abnormal; Resoldering the poorly soldered joint in the circuit block in which abnormality has occurred. Replace IC101 or others as needed.
4-18
PCS-1600/1600P
|
|
 |
> |
|